Medtronic Recallingl its MiniMed 600 Series Insulin Pumps

Informações:

Synopsis

FDA is compelling Medtronic to recall its MiniMed 600 Series Insulin Pumps.  This includes Models 630G and 670G and includes 342,205 devices.  The recall was initiated in November, 2019.  The reason for recalls is a missing or broken retainer ring the helps lock the insulin cartridge into place in the pump’s reservoir compartment.  If the cartridge is not properly locked in place, insulin may be under or over delivered.  There have been 26,421 complaints, 2,175 injuries, and 1 death.  Contact Medtronic Technical Support at 877-585-0166 for more info and device or part replacement.   #diabetes #insulin #insulinpump #recall